
ATI A15/66 Residual Sulfite Monitor
29
O&M Manual
Rev-W (8/10)
ANALOG OUTPUT
A programmable isolated analog output signal is provided for interface with recorders, data
loggers, or controllers. The analog output may be programmed for either 0-20 mA or 4-20 mA output.
This output signal will be linearly proportional to the measured sulfite concentration, and may be scaled
over ranges of 0-0.1 to 0-20.00 PPM. Also, the output may be spanned for non-zero based ranges as
well. Both the zero and full scale output values are programmable so that an output of 1-10 PPM could
be set if desired.
If a zero based voltage output is required, this can easily be accommodated by selecting a 0-20
mA output and placing a shunt resistor across the output. For instance, a 50 ohm resistor will convert the
output to 0-1 VDC while a 250 ohm resistor will provide 0-5 VDC. This will allow the output to
accommodate systems with zero based voltage inputs.
Programming the analog output is done in the CONFIGURATION menu as described in items 11,
12 and 13 on page 24. Connection of the output is made to the terminals shown in Figure 6.
SIMULATE MODE OF OPERATION
As previously noted, A15/66 monitors contain a mode of operation called “SIM” or simulate. This
mode of operation allows the user to adjust the sulfite value manually in order to check devices wired to
either the mA output or the alarm contacts.
To enter simulate mode, press the MODE key until the main display is on the LCD. Press CAL
once and then use the UP key to change the display to SIM. Press ENTER and the display will flash
UPDATE and the display will then read “0.000 SIM A B”.
Pressing the UP arrow will increase the sulfite value on the display. The 4-20 mA output will
increase as the sulfite value increases. Any recorders or remote reading devices that are connected to
the 4-20 mA output should respond to the increasing values. You can adjust the value to any number
desired. If the value you set on the display is above an alarm setpoint, the relay for that setpoint should
activate, as should any alarm or control devices wired to that relay.
Using the simulate mode of operation can be very useful in verifying operation of control or
recording systems that use this equipment. Be sure to return the monitor to AUTO mode of operation
when all simulation work is complete.
TEMPERATURE MEASUREMENT
A temperature element is located in the overflow chamber of the reactor module to measure the
temperature of the sample after stripping. No adjustments are required for this part of the system.
However, it is important to note that the measured temperature is often quite different than the
temperature of the water at the overflow chamber below the monitor. Some sample heating or cooling
can occur in the sample handling system. The temperature measurement is used to compensate the
measurement and must reflect the stripped sample temperature. Do not try to adjust the displayed
temperature to the sample temperature at the overflow. It is not designed to provide this measurement.
There are 2 white wires coming from the sensing element that are connected to the PCB on the terminals
labeled “RTD wiring”. If disconnected make sure the jumper is in place after reassembly.
Kommentare zu diesen Handbüchern