ATI A15/66 Residual Sulfite Monitor
17
O&M Manual
Rev-W (8/10)
START-UP
Prior to operation, recheck electrical connections to be sure that everything is in accordance with
Figures 5 & 6 on pages 12 & 13. If everything is in order, power may be applied to the instrument. When
power is turned on, the display will come up with a program and version number, which will remain on the
display for a few seconds. The display will then change to a display of sulfite concentration and an
indication of the status of control relays. This is the main operating display, and should look like the
display below:
D1 X.XXPPM A B MAIN DISPLAY
The X.XX portion will be numbers. The initial value could be almost anything, depending on
whether the sensor has been connected and the condition of the sensor. The value is not important at
this point. The next portion of the display indicates control relay status. The letters A and B refer to
control relay A and control relay B as marked on the back of the monitor. An open square () indicates a
deactivated relay while a solid square () indicates an activated relay. More information on the control
relays is provided in a later section of this manual.
Beyond the main display, a series of alternate displays are available to the operator. These
displays may be selected by pressing the MODE key on the front panel. It is a good idea to step through
the displays to become familiar with each one. Many of these alternate displays serve as a starting point
for programming functions described later in this manual. The alternate display sequence is as follows:
D2 X.XX PPM SO3 DISPLAY WITHOUT RELAY INDICATION
D3 TEMP: 20.0 C TEMPERATURE DISPLAY
D4 SA X.XXPPM O L RELAY A SETPOINT AND FUNCTION
D5 SB X.XXPPM O L RELAY B SETPOINT AND FUNCTION
D6 CLEANING OFF STATUS OF CLEANING FUNCTION
D7 O1 4.0mA/ 0.00PP ANALOG OUTPUT #1 INDICATION
D8 O2 12.0mA/ 25.0°C ANALOG OUTPUT # 2 INDICATION
D9 CONFIGURATION CONFIGURATION MENU
D10 S/A15 R2.10 PROGRAM AND RELEASE NUMBER