ATI A15/76 Turbidity Monitor Part 1 - Introduction
7
O&M Manual
Rev-H (8/11)
Specifications
Electronics Unit
Range: 4/400 NTU (0-4.000, 0-40.00, 0-400.0)
40/4000 NTU (0-40.00, 0-400.0, 0-4000)
9/999 mg/l (or PPM) SiO
2
(0-9.999, 0-99.99, 0-999.9)
99/9999 mg/l (or PPM) SiO
2
(0-99.99, 0-999.9, 0-9999)
4/400 mg/l PSL (0-4.000, 0-40.00, 0-400.0)
40/4000 mg/l PSL (0-40.00, 0-400.0, 0-4000)
Auto-ranging over 3 decades of concentration
Sensor Type: 90° Scatter (nephelometer)
Accuracy: ± 5% of reading or ± 0.02 NTU, whichever is greater, on 40/400 scale
± 10% of reading or ±2 NTU, whichever is greater, on 400/4000 scale
Linearity: 0.1% of F.S.
Display: 16 character alphanumeric backlit LCD
Control Relays: Two SPDT relays, 5A @ 220 VAC resistive.
Programmable deadband and time delay.
Control Mode: On/Off
Alarm Relay: Independent SPDT relay, 5A @ 220 VAC resistive.
Programmable for actuation on high/low values or control system failure.
Also indicates dry cell condition, fouled sensor, or high ambient light
Analog Output: Isolated 4-20 mA, 600 ohm maximum load. Programmable output range
user set
Operating Conditions: -20° to +55° C., 0-95% R.H. non-condensing.
Power: 110/220 VAC ±10%, 50/60 Hz.
9-36 VDC Optional, 360 mA max. at 12 VDC
Enclosure: Panel mount standard
NEMA 4X wall mount optional
Sensor
Range: 0-4000 NTU
Sensor Type: LED
Measurement angle: 90-degree scatter (nephelometric)
Response Time: 95% in 10 seconds
Power: ± 12 VDC supplied by monitor, preamplifier in sensor head
Temperature Limits: 0 - 50° C.
Pressure Limits: 0 - 100 PSIG
Connections: 7-conductor sensor cable to monitor, 30 feet standard, 350 feet
maximum
Sensor Materials: Delrin body, Acrylic optical windows
Flowcell Materials: PVC
Flowcell Connections: 4 mm tube fittings (black tubing supplied)
Flowcell Pressure: 0-100 PSIG
Flowcell Temperature: 0-60° C